Liquiline System CA80NO
Diagnostics and troubleshooting
Hauser
123
No.
Message
Factory settings
Tests or remedial action
S
D
F
380
Temperature sensor
F
On
On
Temperature sensor defective in reactor, cuvette
or sample pre-heating system
‣
Contact the Service Department and quote the
number displayed.
381
Valve defective
F
On
On
Indicated valve is defective
‣
Contact the Service Department and quote the
number displayed.
385
Heating temp. low
S
On
Off
Sample pre-heating target temperature not
reached
1.
Check fuse.
2.
Check cable and connector.
386
Heating temp. high
S
On
Off
Temperature of sample pre-heating too high
1.
Monitor the temperature of the sample.
2.
Disconnect the power supply if the sample
temperature does not drop.
401
Factory reset
F
On
On
Factory reset is performed
405
Service IP active
C
Off
Off
Service switch is switched on
The device can be addressed at 192.168.1.212.
‣
Switch off the service switch to change to the
saved IP settings.
412
Writing backup
F
On
Off
‣
Wait for the write process to be finished
413
Reading backup
F
On
Off
‣
Wait.
436
SD card
(80%)
M
On
Off
SD card 80% full
1.
Replace SD card with empty card.
2.
Clear SD card.
3.
Set logbook properties to circular buffer
(
Setup
/
General settings
/
Logbooks
).
437
SD card
(100%)
M
On
Off
SD card 100% full. No longer possible to write to
the card.
1.
Replace SD card with empty card.
2.
Clear SD card.
3.
Set logbook properties to circular buffer
(
Setup
/
General settings
/
Logbooks
).
438
SD card removed
M
On
Off
SD card not plugged in
1.
Check SD card.
2.
Replace SD card.
3.
Disable logging.
455
Mathemat. function
F
On
On
Mathematical function: fault condition
1.
Check mathematical function.
2.
Check assigned input variables.
460
Output below limit
S
On
Off
Reasons
• Sensor in air
• Air pockets in assembly
• Sensor fouled
• Incorrect flow to sensor
1.
Check sensor installation.
2.
Clean sensor.
3.
Change assignment of current outputs.
461
Output above limit
S
On
Off
502
No text catalog
F
On
On
‣
Contact the Service Department.