Liquiline System CA80NO
Operation
Hauser
73
Menu/Setup/Sample preparation
Function
Options
Info
Sample preparation 1
In the case of two-channel devices, sample
preparation 2 is also displayed.
Operating mode
Selection
• Controlled
• Independent
Factory setting
Controlled
Controlled:
Sample preparation which is controlled by the
CA80 analyzer,
e.g. Liquiline System CAT820 / CAT860
Independent:
No sample preparation that is controlled by
the Liquiline System CA80 analyzer, e.g.
Stamoclean CAT430. The supply of sample
must be guaranteed by the customer.
If the operating mode selected is
Controlled
:
Tag
Customized text,
32 characters
Cleaning start condition
(CAT860)
Selection
• Immediate
• Date/time
• Disabled
Immediate:
Cleaning starts immediately.
Date/time:
Cleaning starts at the selected date/time.
Disabled:
Cleaning is not active.
If the start condition selected is
Immediate
or
Date/time
:
Cleaning interval
(CAT860)
Selection
• 0-01 to 90-00 (DD-HH)
•
Factory setting
Compressed air or water cleaning for longer
filter maintenance intervals.
Cleaning with liquid cleaner.
If the start condition selected is
Date/time
:
Date
(CAT860)
Selection
• 01.01.1970 to
07.02.2106
DD.MM.YYYY
•
Factory setting
01.01.1970
Time
(CAT860)
Selection
• 00:00:00 to 23:59:59
HH:MM:SS
•
Factory setting
00:00:00
Residence time
(CAT860)
Selection
• 0:30 to 20:00 (MM:SS)
•
Factory setting
1:00
Reaction time of cleaner during the cleaning
Sampling pump interval
(CAT820, CAT860)
Pump time
Selection
10 to 20 s
Factory setting
10 s
Ratio between the pumping duration and the
pumping interval of the peristaltic pump. This
affects the volume of the pumped sample.
Pump pause
Selection
20 to 50 s
Factory setting
30 s
Pressurized air cleaning
(CAT820, CAT860)