Commissioning
Liquiline System CA80FE
38
Hauser
2
3
RK
S1
P
RK
S1
W1
Cu
M
D
S1
RK
D5
D6
D7
7
1
2
W2
optional
dilution
A0033655
25
Liquiline System CA80FE, self-priming
S1
Standard 1
P
Sample
RK
Reagent RK
2, 3, 7
Dispensers
W1
Outlet
D
Outlet
M
Photometer/measuring cell
Cu
Cuvette
1
T-piece
7.1.3
Connecting the sample inlet hose
1. Ensure a constant and sufficient supply of sample at the installation location.
2. Connect the liquid-bearing hoses of the sample supply system.
3. Self-priming system: connect the intake hose supplied (1.5 m (4.92 ft)) to the Liquid
Manager ("sample", → hose connection diagram) and guide it through the hose gland of
the analyzer to the outside.
4. If present, connect the communication cable and hose heater of the sample preparation
system to the analyzer.
5. Ensure that only sample that has a low solids content is supplied, as otherwise there is a
risk of blockage.
7.2
Function check
L
WARNING
Incorrect connection, incorrect supply voltage
Safety risks for staff and device malfunctions!
‣
Check that all connections have been established correctly in accordance with the wiring
diagram.
‣
Ensure that the supply voltage matches the voltage indicated on the nameplate.