Liquiline System CA80FE
Electrical connection
Hauser
27
Only for the "Outdoor" version
10.
Guide the cable of the heated waste hose through the cable gland indicated.
11.
Guide the power supply cable, which is provided by the customer at the installation
location, through the cable gland indicated.
For all versions
12. Route the cables on the rear panel of the device so that they are properly protected. Use
cable clips.
13. Guide the cable to the electronics compartment.
After connecting:
1. Secure the electronics compartment cover with the 6 screws.
2. Fold up the carrier board and use the 6 screws to secure it after connecting.
3. Tighten the cable glands on the bottom of the device to secure the cables.
4. Place the bottle tray back into the housing.
5.2.2
24 V version: connecting the power supply
‣
In the case of devices with a 24 V power supply, the connection cross-section must be at
least 2.5 mm
2
and may not exceed 4 mm
2
.
‣
With 24V power supply, a current of up to 10A can flow. For this reason pay attention to
the voltage drop on the supply line.
‣
The voltage at the device terminals must be within the specified range .