Indumax CLS54
Technical data
Hauser
23
(plus uncertainty of the conductivity of the calibration solution)
9.3
Environment
9.3.1
Ambient temperature range
-20 to +60 °C (-4 to 140 °F)
9.3.2
Storage temperature
-25 to +80 °C (-13 to +176 °F)
9.3.3
Relative humidity
5 to 95 %
9.3.4
Degree of protection
IP 68 / NEMA type 6 (1 m (3.3 ft) water column, 50 °C (122 °F), 168 h)
9.4
Process
9.4.1
Process temperature
-10 to +125 °C (+14 to +257 °F)
9.4.2
Sterilization
150 °C (302 °F) / 6 bar (87 psi) absolute, (max. 60 min.)
9.4.3
Process pressure (absolute)
13 bar (188.5 psi) up to 90 °C (194 °F)
9 bar (130.5 psi) at 125 °C (257 °F)
1 to 6 bar (14.5 to 87 psi) in CRN environment tested with 50 bar (725 psi)
Underpressure down to 0.1 bar (1.45 psi)