Installation
Indumax CLS54
12
Hauser
> 1 m (3.3 f
t)
A0017691
1
Installation positions of the conductivity sensor
If the flow direction changes (after pipe bends), turbulence in the medium can result.
Install the sensor at a distance of at least 1 m (3.3 ft) downstream from a pipe bend.
The product should flow along the hole of the sensor (see the arrows on the housing). The
symmetrical measuring channel allows flow in both directions.
For a 3-A compliant installation, the following must be noted:
After the instrument is installed its hygienic integrity shall be maintained. All process
connections must be 3-A compliant.