Technical data
Indumax CLS54
22
Hauser
8.2
Calibration solutions
Conductivity calibration solutions CLY11
Precision solutions referenced to SRM (Standard Reference Material) by NIST for qualified
calibration of conductivity measuring systems in accordance with ISO 9000
• CLY11-B, 149.6 μS/cm (reference temperature 25 °C (77 °F)), 500 ml (16.9 fl.oz)
Order No. 50081903
• CLY11-C, 1.406 mS/cm (reference temperature 25 °C (77 °F)), 500 ml (16.9 fl.oz)
Order No. 50081904
• CLY11-D, 12.64 mS/cm (reference temperature 25 °C (77 °F)), 500 ml (16.9 fl.oz)
Order No. 50081905
• CLY11-E, 107.00 mS/cm (reference temperature 25 °C (77 °F)), 500 ml (16.9 fl.oz)
Order No. 50081906
Technical Information TI00162C
9
Technical data
9.1
Input
9.1.1
Measured values
• Conductivity
• Temperature
9.1.2
Measuring range
Conductivity
Recommended range: 100 μS/cm to 2000 mS/cm
(uncompensated)
Temperature
–10 to +150 °C (+14 to +302 °F)
9.1.3
Cell constant
k = 6.3 cm
–1
9.1.4
Temperature measurement
Pt1000 (according to DIN EN 60751)
9.2
Performance characteristics
9.2.1
Temperature response time
t
90
≤ 26 s
9.2.2
Maximum measured error
± (0.5 % of r 10 μS/cm) after calibration