Indumax CLS54
Electrical connection
Hauser
19
5.1.2
Sensors with CSA approval
The instructions in the Control Drawing apply for sensors with FM or CSA approval. You can
find the Control Drawing in the Operating Instructions of the transmitter used.
5.2
Connecting to the transmitter
The sensor is supplied with a fixed cable. The wiring diagram is provided in the Operating
Instructions of the transmitter used.
Connection via a VBM junction box is necessary for a cable connection. The extension to the
transmitter is via the CLK6 cable.
WH
RD
Pt 1000
YE
GN
Ferrule RD
WH
Ferrule BU
Screen
Inner connector
BN n.c.
Cable
}
Ferrule RD
Ferrule BU
Screen
Inner connector
Coax WH
}
}
Coax RD
A0005433-EN
8
Fixed cable / measuring cable CLK6
Cable length:
maximum 55 m (180 ft) total length (non-Ex
versions)
maximum 50 m (180 ft) total length (Ex versions)
5.3
Ensuring the degree of protection
Only the mechanical and electrical connections which are described in these instructions and
which are necessary for the required, designated use, may be carried out on the device
delivered.
‣
Exercise care when carrying out the work.
Otherwise, the individual types of protection (Ingress Protection (IP), electrical safety, EMC
interference immunity) agreed for this product can no longer be guaranteed due, for example,
to covers being left off or cable (ends) which are loose or insufficiently secured.