Commissioning
Proline Promag 51
58
Hauser
6.5
Configuration
6.5.1
Current output: active/passive
The current output is configured as "active" or "passive" by means of various jumpers on the
I/O board.
#
Warning!
Risk of electric shock. Exposed components carry dangerous voltages. Make sure that the power
supply is switched off before you remove the cover of the electronics compartment.
1.
Switch off power supply.
2.
74.
3.
Position the jumpers
"
Caution!
Risk of destroying the measuring device. Set the jumpers exactly as shown in the graphic.
Incorrectly set jumpers can cause overcurrents that would destroy either the measuring device
or external devices connected to it.
4.
Installation of the I/O board is the reverse of the removal procedure.
A0001044
Fig. 44:
Configuring the current outputs with the jumpers (I/O board)
1
Active current output (factory setting)
2
Passive current output
1
2