Rackbus
For system use, wire the Rackbus as shown in Fig. 2.15.
•
For ZA 672 (Modbus) - Computer/PLC wiring see Operating Instructions
BA 054E
for ZA 673 (PROFIBUS) see Operating Instructions BA 073E
for ZA 674 (FIP) see Operating Instructions BA 085E
•
Be sure that the system is properly grounded. A difference in ground potential
between the network device and Gateway can cause loss of data or
transmission faults
Relays
Alarms and limit switch relays can be connected as shown in Fig. 2.16.
•
Max. current per contact 2.5A : for switching capacity see Section 2.6.
Alternative 2
+
RACKBUS
d b z
2
4
6
10 +
8
12 + +
14
16
32 L+
+
RACKBUS
d b z
2
4
6
10 +
8
12 + +
14
16
32 L+
+
RACKBUS
d b z
10 +
18 r
20 u a
22 r
24 u a
26 r
28 u a
30 L-
18 r
20 u a
22 r
24 u a
26 r
28 u a
30 L-
L+
L-
2
4
6
8
14
16
18
20
22
28
26
24
30
32
12
u
a
r
u
a
r
u
a
r
BA082E17
Max. 64
Commutec-
transmitters in
parallel on
Rackbus
Commutec transmitters FTG, FMB, FMC, FMU, PMX, TMT etc.
ZA 672
FTG 671
Fig. 2.15:
Rackbus wiring diagram
Max.current
per line
2,5 A
BA0082E18
Recommended
Alternative 1
Fig. 2.16:
Suggestions for wiring
together relays and alarms
Gammapilot FTG 671
Chapter 2: Installation
22
Summary of Contents for gammapilot FTG 671
Page 4: ...Gammapilot FTG 671 Table of Contents 2...
Page 12: ...Gammapilot FTG 671 Chapter 1 Introduction 10...
Page 28: ...Gammapilot FTG 671 Chapter 2 Installation 26...
Page 32: ...Gammapilot FTG 671 Chapter 3 Controls 30...
Page 48: ...Gammapilot FTG 671 Chapter 5 Limit Switches Relays 46...
Page 50: ...Gammapilot FTG 671 Chapter 6 Maintenance 48...
Page 59: ...Gammapilot FTG 671 Operating Matrix 57...
Page 61: ...Index Gammapilot FTG 671 59...
Page 62: ...Gammapilot FTG 671 Index 60...
Page 63: ...Index Gammapilot FTG 671 61...