Activity of the source
Hauser radiometric systems operating with the DG 57 rod scintillation counters
utilize the lowest source strengths.
For limit switching, the source strength is calculated such that the dose rate arriving at
the detector is:
•
approx. 0.1
…
1
µ
Sv/h (0.01
…
0.1 mR/h) for DG 57 detector
•
approx. 1.2
…
8
µ
Sv/h (0.12
…
0.8 mR/h) for DG 17 Z detector
•
approx. 0.6
…
4
µ
Sv/h (0.06
…
0.4 mR/h) for DG 27 Z detector.
In this case, the dose rate usually lies below that defining a control area, i.e. 7.5
µ
Sv/h
(0.75 mR/h) or 2.5
µ
Sv/h (0.25 mR/h), which has to be sealed off.
Warning!
•
The actual extent of the control area depends on local regulations (other limits may
apply) and must be determined on the basis of dose rate measurements made on
site on both sides of the measuring system prior to the commissioning of the
equipment.
•
If the front of the source container can be accessed, shielding or other protective
measures must be provided to prevent accidental exposure to the beam!
Gammapilot FTG 671
Chapter 2: Installation
14
Summary of Contents for gammapilot FTG 671
Page 4: ...Gammapilot FTG 671 Table of Contents 2...
Page 12: ...Gammapilot FTG 671 Chapter 1 Introduction 10...
Page 28: ...Gammapilot FTG 671 Chapter 2 Installation 26...
Page 32: ...Gammapilot FTG 671 Chapter 3 Controls 30...
Page 48: ...Gammapilot FTG 671 Chapter 5 Limit Switches Relays 46...
Page 50: ...Gammapilot FTG 671 Chapter 6 Maintenance 48...
Page 59: ...Gammapilot FTG 671 Operating Matrix 57...
Page 61: ...Index Gammapilot FTG 671 59...
Page 62: ...Gammapilot FTG 671 Index 60...
Page 63: ...Index Gammapilot FTG 671 61...