Emergency procedure
Source container FQG66
36
Hauser
9
Emergency procedure
9.1
Objective and overview
This emergency procedure shall be put into effect immediately to secure an area in the
interests of protecting personnel where an exposed source is known, or suspected, to exist.
Such an emergency exists when a radioisotope is exposed either by it becoming separated
from the source container or a source container cannot be put into the "AUS/OFF" position.
This procedure will safeguard the personnel until the responsible radiation safety officer
will attend site and advise on corrective action.
The custodian of the radioactive source (the customer's designated "authorized person") is
responsible for observing this procedure.
9.2
Emergency procedure
1. Determine the unsafe area by on-site measurement.
2. Cordon off the concerned area by yellow tape or rope and post international radiation
warning signs.
9.2.1
The shutter cannot be switched to the "AUS/OFF" position
In this case the radiation source container should be unbolted from its mounting position
and
• in case the radiation emission channel is located horizontally: put the emission channel
towards a very thick wall, or
• in case the radiation emission channel is located vertically: put the source container on
its device base on the ground
L
CAUTION
Risk of injury
‣
Use lifting tools due to the weight of approx. 435 kg (959.18 lb).
In case of a shutter that will not close due to failure of the pneumatic drive:
Switch off or disconnect compressed air, remove supply lines if necessary. Remove the
protection cap. Unscrew the connection screw between the eye bolt at the end of the
source holder rod and the pneumatic drive. Move the source holder rod to "AUS/OFF"
position and secure it by the means of the locking bolt.
1
2
A0027945
1
Eye bolt
2
Connection screw