Cerabar S 4 to 20mA HART
Description of parameters
Hauser
31
P01-MIDxxxxx-19-xx-xx-EN-015
Fig. 13:
OPERATING and DIAGNOSTICS group
For the OPERATING group, see Page 31, Table 13
For the SIMULATION MODE function group, see Page 33, Table 14
For the MESSAGES function group, see Page 33, Table 15
For the USER LIMITS function group, see Page 35, Table 16
*
I
J
*
I
J
J
J
J
J
1)
1)
2)
*
*
none
Pressure
Tank content
Current
Alarm Warning
LANGUAGE
SETTINGS
POSITION ADJUSTMENT
OPERATING MENU
GROUP SELECTION
Measured value
SIM. PRESSURE
SIM. TANK CONT.
SIM. CURRENT
SIM. ERROR NO.
OPERATION
ENTER RESET CODE
OPERATING HOURS
INSERT PIN NO
DIAGNOSTICS
SIMULATION
SIMULATION MODE
MESSAGES
RESET ALL ALARMS
SELECT ALARMTYPE
ALARM DELAY
ALARM DISPL. TIME
ALARM STATUS
LAST DIAG. CODE
ACK. ALARM MODE
ACK. ALARM
Pmin ALARM WINDOW
Pmax ALARM WINDOW
Tmin ALARM WINDOW
Tmax ALARM WINDOW
USER LIMITS
HistoROM AVAIL.
HistoROM CONTROL
ERROR No.
Parameters marked with an asterisk "*"
are not always displayed. The display
depends on the settings of other parameters,
see the following Table.
1) Display via on-site display only
2) Display via digital communication
DOWNLOAD SELECT.
Table 13: (GROUP SELECTION
) OPERATING MENU
OPERATING
Parameter name
Description
ENTER RESET CODE (047)
Entry
Reset parameters completely or partially to factory values or delivery status.
See also Operating Instructions BA00412P, Section "Factory setting (reset)".
Factory setting:
0
OPERATING HOURS (409)
Display
Displays the hours of operation. This parameter cannot be reset.