System integration
Cerabar PMP23 IO-Link
22
Hauser
Designation
ISDU
(dec)
ISDU
(hex)
Size
(byte)
Data
type
Access Default
value
Value range
Offset /
Gradient
Data
storage
Range
limits
Output 1 (Ou1)
81
0x0051
1
unit
r/w
HNO
0 ~ HNO
1)
1 ~ HNC
1)
2 ~ FNO
1)
3 ~ FNC
1)
Yes
Hi Max value (maximum
indicator)
82
0x0052
4
int
ro
No
Lo Min value (minimum
indicator)
83
0x0053
4
int
ro
No
Revisioncounter (RVC)
84
0x0054
2
unit
ro
No
1)
For an explanation of the abbreviations, see the parameter description → 32
8.2.2
IO-Link-specific device data
Designation
ISDU (dec)
ISDU (hex)
Size (byte)
Data type
Access Default value
Data storage
Serial number
21
0x0015
max. 16
String
ro
Firmware version
23
0x0017
max. 64
String
ro
ProductID
19
0x0013
max. 64
String
ro
PMP23
ProductName
18
0x0012
max. 64
String
ro
Cerabar
ProductText
20
0x0014
max. 64
String
ro
Absolute and gauge pressure
VendorName
16
0x0010
max. 64
String
ro
Hauser
VendorText
17
0x0011
max. 64
String
ro
People for Process Automation
Hardware Revision
22
0x0016
max. 64
String
ro
Application Specific Tag
24
0x0018
32
String
r/w
Device Type
256
0x0100
2
Uinteger16
ro
0x92FF
Actual Diagnostics (STA)
260
0x0104
4
String
ro
No
Last Diagnostic (LST)
261
0x0105
4
String
ro
No
8.2.3
System commands
Designation
ISDU (dec)
ISDU (hex)
Value range
Access
Reset to factory settings (RES)
2
0x0002
130
w
Device Access Locks.Data Storage Lock
12
0x000C
0 ~ False
2 ~ True
rw
8.3
Overview of diagnostic events
→ 38