PlasmaQuant PQ 9000 (Elite)
Installation and commissioning
43
Figure 22 In-line filter in sampling compartment
1 Sample hose to nebulizer
2 Filter block with hose connectors
3 Sample pump hose
1.
Insert the in-line filter into the hose path such that the arrow on the filter block
points in the direction of flow (i.e. in the direction of the nebulizer).
2.
Assemble the in-line filter as described on the enclosed data sheet.
To do so, insert the nebulizer hose into the banjo bolt using a ferrule. The conical
side of the ferrule must point toward the banjo bolt.
1 Ferrule
2 Banjo bolt
3 Hose
Figure 23 Position of the ferrule in the banjo bolt
Screw the hose connector into the outlet of the filter block.
Screw a short capillary hose into the inlet of the filter block using another hose con-
nector.
3.
Connect the capillary hose with the pump hose for the sample. To do so, push the
capillary hose into the pump hose.
4.
Connect the nebulizer hose with the nebulizer.
Installing the in-line filter