PlasmaQuant PQ 9000 (Elite)
Installation and commissioning
41
Figure 20 Isomist in the sampling compartment
1 Mains cable
2 USB port
3 Waste hose
4 Sample hose on the nebulizer
5 Nebulizer
6 Isomist
7 Carrier gas hose
8 Transfer tube
4.6.2
Installing the argon humidifier
The argon humidifier is suitable for the analysis of samples with a high salt content. By
humidifying the carrier gas the argon humidifier prevents that salts crystallize inside the
nebulizer or injector, blocking it. In addition, it improves the stability of the signal and
the recovery.
The carrier gas, argon, flows via a membrane coil through deionized water and is satu-
rated with water vapor on its way. With the help of the bypass valve (3 in Figure 22),
the argon humidification can be switched on or off without needing to disconnect the
hose lines.
Please refer to the enclosed data sheet for a description of the setup and maintenance
of the argon humidifier.