Rev B 2022/03
107
CyBio Well vario
Maintenance & Care
9.2.6
Cleaning the Carriage (Linear Transporting Unit)
To clean the carriage, proceed as follows:
1.
Switch CyBio Well vario power off at the main power switch, then disconnect the
power plug from the socket.
2.
Remove tip magazine from pipetting head for safety reasons (
).
3.
Remove all microplates, reservoirs or washing troughs.
4.
Remove carriage from the transport rails by raising it at the front and the back end by
equal amounts simultaneously. Proceed with due care, in order to prevent damage.
5.
Clean the carriage and transport rails using a mild detergent or disinfectant.
6.
Carefully place the carriage back onto the transport rails.
7.
Carefully push tip magazine back into the holder at the pipetting head. Make sure
that tip magazine is correctly placed on completion of work.
8.
Reconnect power plug to mains socket and switch CyBio Well vario power on at the
main switch.
The system will be re-initialized and becomes operational again.
9.2.7
Cleaning the Rotary Stage (Circular Transporting Unit)
To clean the rotary stage, proceed as follows:
1.
Switch CyBio Well vario power off at the main power switch, then disconnect the
power plug from the socket.
2.
Remove tip magazine from pipetting head for safety reasons (
).
3.
Remove all microplates, reservoirs or washing troughs.
4.
Clean the rotary stage using a mild detergent or disinfectant.
5.
Carefully push tip magazine back into the holder at the pipetting head. Make sure
that tip magazine is correctly placed on completion of work.
6.
Reconnect power plug to mains socket and switch CyBio Well vario power on at the
main switch.
The system will be re-initialized and becomes operational again.
9.2.8
Sealing Plate (Pipetting Head)
You are advised to follow these rules:
1.
Do not use pipette tips covered with dust.
2.
Check sealing plate for cleanness once in a month.
3.
Trigger a tip changing cycle and remove the tip magazine for inspection and cleaning.
The sealing plate can now be accessed from the bottom. Use a non-shedding slightly
moistened cloth for cleaning.
4.
Proceed with utmost caution in order to prevent unwanted position-shifting of the
sealing plate. Otherwise, there is danger of plate openings getting clogged with fluff
balls.
TIP
Carriage performance depends on the current state of the transport rails. Remove pock-
ets of dirt, contamination or damage of transport rails immediately.
TIP
Where pipette tips are changed at shorter intervals, fine fluff-ball stock or dust may set-
tle on the sealing plate, thus impairing the efficiency of sealing.
Summary of Contents for Analytik Jena CyBio Well vario
Page 1: ...User Manual CyBio Well vario ...
Page 3: ...Rev B 2022 03 III CyBio Well vario ...
Page 5: ...Rev B 2022 03 V CyBio Well vario ...
Page 6: ...VI Rev B 2022 03 CyBio Well vario ...
Page 7: ...Rev B 2022 03 VII CyBio Well vario ...
Page 8: ...VIII Rev B 2022 03 CyBio Well vario ...
Page 12: ...XII Rev B 2022 03 CyBio Well vario ...
Page 16: ...XVI Rev B 2022 03 CyBio Well vario ...
Page 26: ...10 Rev B 2022 03 Technical Specifications CyBio Well vario ...
Page 62: ...46 Rev B 2022 03 Technical Description CyBio Well vario ...
Page 118: ...102 Rev B 2022 03 Troubleshooting CyBio Well vario ...
Page 126: ...110 Rev B 2022 03 Decommissioning CyBio Well vario ...
Page 130: ...114 Rev B 2022 03 Accessories Spare Parts CyBio Well vario ...
Page 132: ...116 Rev B 2022 03 Waste Disposal CyBio Well vario ...
Page 142: ...126 Rev B 2022 03 CyBio Well vario ...