Rev B 2022/03
93
CyBio Well vario
Operation
7.4.3
Volume Cycle
Standard cycle:
Aspiration with additional stroke
Dispensing (single volume or subvolumes)
Residual dispensing
Motion to zero-position
Simplified cycle (only for dispensing
of single volumes):
Aspiration without additional stroke
Dispensing with additional stroke
Motion to zero-position
7.4.4
Waiting Time
Piston motion always creates a state of underpressure or overpressure during an aspiration
or a dispensing cycle. In order to maintain the specified system precision, pressure equal-
ization in the tips must be allowed to finish. This is supported by a specific "Interruption"
software command. A waiting time of two seconds proves long enough with most watery
solutions.
Summary of Contents for Analytik Jena CyBio Well vario
Page 1: ...User Manual CyBio Well vario ...
Page 3: ...Rev B 2022 03 III CyBio Well vario ...
Page 5: ...Rev B 2022 03 V CyBio Well vario ...
Page 6: ...VI Rev B 2022 03 CyBio Well vario ...
Page 7: ...Rev B 2022 03 VII CyBio Well vario ...
Page 8: ...VIII Rev B 2022 03 CyBio Well vario ...
Page 12: ...XII Rev B 2022 03 CyBio Well vario ...
Page 16: ...XVI Rev B 2022 03 CyBio Well vario ...
Page 26: ...10 Rev B 2022 03 Technical Specifications CyBio Well vario ...
Page 62: ...46 Rev B 2022 03 Technical Description CyBio Well vario ...
Page 118: ...102 Rev B 2022 03 Troubleshooting CyBio Well vario ...
Page 126: ...110 Rev B 2022 03 Decommissioning CyBio Well vario ...
Page 130: ...114 Rev B 2022 03 Accessories Spare Parts CyBio Well vario ...
Page 132: ...116 Rev B 2022 03 Waste Disposal CyBio Well vario ...
Page 142: ...126 Rev B 2022 03 CyBio Well vario ...