72
Rev B 2022/03
Operation
CyBio Well vario
5.
Confirm your selection by pressing
, in order to trigger a dispensing cycle.
The previously aspirated volume is dispensed again. While dispensing lasts, the
lifter and carriage (rotary stage) control unit remains inactive.
Tip replacement
1.
Select tip change basic function in
[Manual]
menu.
The magazine plate with capillary tips is automatically released.
Fig. 78:
Display – tip change – tip release
2.
Remove tip magazine from the pipettor.
Fig. 79:
Display – tip change – magazine removal
Fig. 80:
Capillary magazine change (version: head replacement from the front)
CAUTION
There is danger of finger pinching and damage to capillary tips!
The magazine plate is released and clamped with electro-motoric power. For a tip
change, the magazine plates must be inserted until hard stop position. Prevent physical
contact with the magazine plate as automatic tip clamping is in process!
1 Capillary magazine
2 Capillary tips
Changing tips
Loosen tips
Changing tips
Remove tip magazine
1
2
Summary of Contents for Analytik Jena CyBio Well vario
Page 1: ...User Manual CyBio Well vario ...
Page 3: ...Rev B 2022 03 III CyBio Well vario ...
Page 5: ...Rev B 2022 03 V CyBio Well vario ...
Page 6: ...VI Rev B 2022 03 CyBio Well vario ...
Page 7: ...Rev B 2022 03 VII CyBio Well vario ...
Page 8: ...VIII Rev B 2022 03 CyBio Well vario ...
Page 12: ...XII Rev B 2022 03 CyBio Well vario ...
Page 16: ...XVI Rev B 2022 03 CyBio Well vario ...
Page 26: ...10 Rev B 2022 03 Technical Specifications CyBio Well vario ...
Page 62: ...46 Rev B 2022 03 Technical Description CyBio Well vario ...
Page 118: ...102 Rev B 2022 03 Troubleshooting CyBio Well vario ...
Page 126: ...110 Rev B 2022 03 Decommissioning CyBio Well vario ...
Page 130: ...114 Rev B 2022 03 Accessories Spare Parts CyBio Well vario ...
Page 132: ...116 Rev B 2022 03 Waste Disposal CyBio Well vario ...
Page 142: ...126 Rev B 2022 03 CyBio Well vario ...