Eccentric Cam
___________________________________
Important: Before attempting to remove the cover, switch off mains
and remove power supply cable from rear of sieve shaker.
In order to adjust the gap on the eccentric cam remove the cover as
follows:
1.
Remove the two clamp rods.
2.
Remove the four cover screws and the blanking plug.
3.
Remove the mains inlet connector by unscrewing the two M3
countersunk screws securing the inlet connector and draw out
to clear the cover.
4.
Remove the mains switch.
5.
The switch is a snap fitting and can be gently levered out, do
not disconnect leads.
6.
Lift up the cover assembly, taking care to ease up the rear to
clear the electrical components at the control panel over the
location casting.
7.
Disconnect the wiring to the motor and the earth lead.
8.
The cover assembly may now be completely removed.
9.
To reassemble the shaker, reverse the order of 1 to 8 above.