1. Position the chair on a flat surface free from any obstruction. Always allow sufficient
space behind and in front of the chair to allow the chair to be fully reclined (See figure
3). Do not allow the chair to touch the wall or other furniture when in use.
2. Apply the castor brakes (See Page 18).
3. If the Battery Operated Reset System is to be used, install the batteries (see Battery
Operated Reset System on page 27).
4. Plug the power cable into a power socket and switch the socket on.
5. Adjust the chair to an upright position (See ‘Hand Controls’ and ‘Changing Chair
Positions’) on Pages 20 and 21 respectively.
6. Check that the backrest fully reclines and returns to the upright position without being
obstructed.
7. Tilt the chair to its maximum and return to the riser position.
8. Adjust the chair again to the upright position.
9. Check that the upholstery is not damaged and is not showing signs of excessive wear.
6. Before Use Checks
Always apply the castor brakes before seating a user in the
chair.
WARNING
!
Before use, always do the following:
The chair is ready for use if all the movements are smooth and no problems are detected.
1702 mm (67")
400 mm (15.5")
650 mm (25.5")
Figure 3 - Chair to wall / furniture measurement