It is very important to check the calibration setup for correctness
- Cabling
- Ground reference is conductive
- CDN or clamp is correct connected
The CAL data of cal store can only be lost due to a discharged internal battery or by overwriting existing data. In
case of battery discharge, the calibration stores are cleared and marked invalid.
Calibration
Which calibration store should be overwritten?
F1 : Calibration Store F1
invalid
F2 : Calibration Store F2
invalid
F3 : Calibration Store F3
invalid
F4 : Calibration Store F4
invalid
F5 : Calibration Store F5
invalid
F1
F2
F3
F4
F5
F6
F7
A nearly unlimited number of CAL data can be generated, managed and included in the actual operation of the
CWS 500N1.x
. If the operator is working with more than two CDN’s or clamps, the internal stores F1 / F1 are
not sufficient and ICD will have to be used which assures the security of data storage.
Fig 5.8: Calibration data CDN
Fig 5.9: Transfer Characteristic EM Clamp