5.5.1. Calibration
Menu Calibration
Calibration Procedure
Calibration
Calibration
Which calibration - store should
be overwritten?
Calibration - Store F2
F1 : Calibration Store F1
valid
Will be overwritten
F2 : Calibration Store F2
valid
F3 : Calibration Store F3
valid
F4 : Calibration Store F4
valid
F5 : Calibration Store F5
invalid
Start
F1
F2
F3
F4
F5
F6
F7
F1
F2
F3
F4
F5
F6
F7
After selecting one of the calibration stores the user may start the calibration. The firmware will ask whether to
overwrite the existing calibration store or not.
A new CAL routine will overwrite the selected store F1 / F5. The operator has to confirm that he wants to
overwrite the existing data.
Calibration
Calibration Store F2
Calibration in Progress
Break
f = 000.100 MHz
F1
F2
F3
F4
F5
F6
F7
During the calibration the display shows the actual calibration frequency
By pressing BREAK, the calibration will be stopped and the firmware will restore the calibration data from the
previous calibration.
If the CWS 500N1.x is unable to complete the calibration procedure (out of range, calibration setup incorrect or
incomplete, etc.) the firmware will restore the calibration data from the previous calibration.
Typical calibration data files are stored in the software icd.control