EM TEST (Switzerland) GmbH
CWS500 N1.3 / N1.4
User manual
V 1.01
14 / 47
5.
Manual Operation
5.1.
Basic Operations
The simulator operates by an easy menu control system. Seven
function keys are available to select parameters and functions. All
functions are indicated on the display.
Fig 5.1
The selected parameter is blinking and can be changed by turning the knob (incr./ decrease.).
:
Select the digit to change with the cursor (
).
ESC
:
ESC will take you back to the previous level in the menu and set the displayed
values. The latest settings are stored automatically and will be recalled when the
menu is selected again.
EXIT
:
The firmware will reset to the main screen.
Fig 5.2
Start-up display
CWS 500N1.3
Continuous Wave Simulator
RF 9 kHz
– 1 GHz
V 8.50
SWN: 001234
Generator Model
Frequency Ranges
The version number and the software number SWN used for tracing purposes. The report and calibration
certificates list these numbers. Even the software report generator lists these numbers.
Page 1 (Main Menu)
MAIN MENU
F1 : Test Routines
F2 : Calibration and Saturation Test
F3 : Freestyle mode 9kHz
– 1 GHz
F7 : Service
F1
F2
F3
F4
F5
F6
F7
F1 Test Routines
The firmware includes the test routines for IEC 61000-4-6 standard.
F2 Calibration and Saturation Test
This menu includes calibration routines for the test. The amplifier saturation-check routine with 5.1dB higher
level, than calibrated, completes the Ed. 4 requirement. The functions are the same as selecting Cal within a
test routine.
F3 Freestyle Mode 9 kHz
– 1 GHz
This menu operates the internal signal generator manually. For frequencies from 400 MHz up to 1 GHz (N1.3),
or 9 kHz-100 kHz and 300 MHz
– 1 GHz ( N4.x), external amplifier is used.
F7 Service
Select and displays the setup.