EM TEST (Switzerland) GmbH
CWS500 N1.3 / N1.4
User manual
V 1.01
19 / 47
5.5.
Calibration and Saturation test Menu
After selecting CAL in any of the internal programs of the CWS 500N1.x or selecting the
“Calibration and
Saturation Test” in the main menu, the display will show the calibration stores.
Structure of the calibration Menu
MAIN MENU
F1 : Test Routines
F2 : Calibration and Saturation Test
F3 : Freestyle mode 9kHz – 1 GHz
F7 : Service
Calibration and Amplifier Saturation-Test
F1 : Calibration
F2 : Amplifier Saturation Test
Calibration
Which Calibration Stroe should
be overwritten?
F1 : Calibration Store F1 valid
F2 : Calibration Store F2 valid
F3 : Calibration Store F3 valid
F4 : Calibration Store F4 valid
F5 : Calibration Store F5 valid
Calibration
Calibration Store F1 .. F5
will be overwritten
Start
Kalibration
Calibration Store F1 … F5
Calibration in Progress
Fail1 Kalibration stopped; Fail2 Calibration interrupted;
Calibration finished; Kalibration out of range; Calibration
interrupted
Break
f=xxx KHz
Amplifier Saturation-Test
U = 10.0 V
Cal = F1
START AENDERN
Amplifier Saturation-Test
U = 10.0 V Calibration Store F1
Amplifier Saturation-Test running
Fail1 Saturation-Test stopped; Fail2 Saturation-Test
stopped; Amplifier Saturation -Test finidhed; Amplifier
Saturation-Test interrupted; Amplifier Saturation-Test Fail
Break fwd = x.xx V; f=xxx KHz
Calibration / Amplifier Saturation-Test
…
Pass / Fail
F2
F2
F1
...
F5
F1
F1
F1 /
Test End
F1 /
Test End
Amplifier Saturation-Test
No valid calibration found
Please start first a calibration
OK
Amplifier Saturation-Test
U
Cal
10
F1
F1
F2
F1
ESC
EXIT
MAIN MENU
Return to last menu