Warnings or Cautions
Light Warning
or Caution
Cause
Corrective Action
Yellow light con-
stant with beep
sequence every
30 seconds
Low Battery: Controller
not recently recharged.
Press OFF to terminate the
treatment. Place the Control-
ler in a Charging Station until
fully charged.
Red light and
selected green
light flashing
rapidly with beep
sequence every
30 seconds
Current Interrupt: May
be caused by a loose
connection, insuf-
ficiently hydrated drug
pads or poor contact
between the Patch and
the skin.
Press PAUSE to pause the
treatment. Check for loose
connections, properly hy-
drated drug pads and good
contact between the Patch
and the skin. After correcting
the problem, press START to
continue the treatment.
Automatic Time Calculation -
All time calculations are performed automatically, even if the current
setting is changed.
Automatic Current Ramp Up -
When the start button is pushed, the Controller automatically ramps up
the current output to the desired setting. While the current is ramping,
the green light will flash rapidly. When the current reaches the setting
selected, the green light will be on constantly.
7