Automatic Current Ramp Down -
After reaching the preset dose, the current automatically decreases,
returning to 0 mA, indicating the treatment is complete. Also, automatic
current ramp down takes place if the ON/OFF button is pressed (i.e.
Treatment Abort) or if the PAUSE button is pressed.
Automatic Fast Current Shut Off -
If there is a Low Battery Alarm or Current Interruption Alarm, the
current will be immediately set to 0 mA.
PATCH DESCRIPTION
The Hybresis Patch is a disposable, single-use Patch with an internal
battery and current-limiting circuitry. It can deliver both negatively and
positively charged water-soluble drugs. The Patch-Only Treatment can
be used with and without the Controller. The Controller is used with the
Patch to deliver a Hybresis or Standard Treatment. A Patch Treatment
does not require the use of the Controller.
Drug Polarity Labeling -
The polarity of the drug pads are labeled on the Patch.
Patch Drug Pads -
The Patch has two drug pads. Each drug pad has a ~1.5 mL fill volume.
Use negatively charged water-soluble drugs on the negative (-) drug pad
and positively charged water-soluble drugs on the positive (+) drug pad.
Battery Pack and Controller Connector -
The Battery Pack that contains two 3-Volt batteries is also the
connecting location for the Controller.
8