Standard Treatment
The Controller delivers current at 2, 3 or 4 mA to the Patch for 10-20
minutes, resulting in a 40 mA-minute treatment. For an 80 mA-minute
treatment, repeat the treatment.
Patch-Only Treatment
The Patch delivers low level current over 2-4 hours, resulting in an
approximate 40-80 mA-minute treatment respectively.
HYBRESIS MODE TREATMENT
NOTE: While using the Controller, should an in-process iontophoresis
treatment need to be stopped or paused, DO NOT suddenly remove
the Controller from the Patch without first switching off the Controller.
To stop a treatment while the Controller is administering iontophoresis,
press the ON/OFF button and wait a few moments for the Controller
to turn off.
1. Examine the Controller before use to assure that the two prongs that
insert into the Patch are free of grime and debris. Dirty contacts could
cause erratic behavior during treatment. If needed, the prongs may be
cleaned with an alcohol wipe.
2. Push the ON button on the Controller. The Green Hybresis Light will
blink slowly.
3. Attach the Controller to the Patch. The Patch connector (located at the
center of the Patch) plugs into the slot on the back of the Controller.
Ensure that the Patch connector is fully and securely
engaged into the Controller- a click will be heard upon full engagement.
4. Position the patient so that there is no pressure on the Patch during
treatment.
5. Press the START button to begin treatment. The Green Hybresis Light
will blink more rapidly, then glow steadily.
12