13
6. After three minutes, the Controller will sound a beep and the
lights will turn off automatically. This indicates that the Skin
Conductivity Enhancement (SCE) is completed.
7. Remove the Controller from the Patch. The Patch will now continue
to deliver the remainder of the iontophoresis treatment to the
patient.
8. The average time to complete the dose is indicated in the following
table. To prevent excessive dosing, the Patch automatically
switches off iontophoresis after the maximum dose has been
administered.
9. Instruct the patient to remove and discard the Patch after a
minimum of one to two hours for a 40 to 80 mA-minute dose
respectively.
10. Discard the Patch after treatment has been completed. The Patch
cannot be reused.
STANDARD MODE TREATMENT
NOTE: While using the Controller, should an in-process iontophoresis
treatment need to be stopped or paused, DO NOT suddenly remove
the Controller from the Patch without first switching off the Controller.
To stop or pause a treatment while the Controller is administering
iontophoresis, press the ON/OFF button and wait a few moments for
the Controller to turn off.
1. Press the ON/OFF button on the Controller. The Green Hybresis
Light will blink slowly.
.
Hybresis Mod
e
40 mA-minutes
60 mA-minutes
80 mA-minutes
Wear Time
1 hour
1.5 hours
2 hours