Charging
The single internal 3.7 VDC rechargeable lithium-ion battery cell
is connected to a Protection Circuit Module (PCM) that provides
additional protection for over-charging and over-discharging. However,
the battery will enter an over-discharge state via self-discharge if it
is not used for a long time. This results in battery degradation (e.g.
the mA-Hr capacity of the battery will decrease). To prevent over-
discharging, return the Controller to the Charging Station after each
use. If the Controller cannot be stored in the Charging Station for a
considerable period of time (e.g. 6 months), fully charge the Controller
before long-term storage.
Lithium-ion batteries, unlike Ni-Cad batteries, do not suffer from
“Memory Effect” (the lithium-ion battery does not require a periodic full
discharge and recharge to “refresh” the battery). The Controller can be
stored and recharged in the Charging Station after each use.
Elevated temperature will result in degraded battery performance and
reduced battery service life. Therefore, never place the Controller near
heating sources or in direct sunlight for long periods. DO NOT open the
Controller case to access the battery or for any other reason.
Service
For clinical questions, contact the Professional Services Department at
1-800-328-2536 ext. 8506. If any component of the Hybresis System is
not functioning properly or requires servicing, contact the Empi Repair
Department at 1-800-862-2343.