9.3.13
Broken Glass
The sensing element in the pH sensor is a thin glass membrane. Normally, the impedance
of the glass membrane is about 80 - 100 MΩ. If the glass membrane gets broken or
cracked, the impedance drops to less than 10 MΩ.
Procedure
1. Check sensor settings under the
Measurement
submenu. Confirm that the
preamplifier location is set to transmitter.
2. Confirm that the pH sensor is installed in the flow cell and sample is flowing through
the cell.
3. Check the sensor response in two buffers having different pH values.
If the membrane is cracked or broken, the pH reading will be about the same in
both buffers.
4. Replace the pH sensor.
9.4
Troubleshooting when a Warning message is
showing
Warning message
Explanation
Section
Sensor No solution Gnd
pH sensor may be miswired.
Sensor Need Factory Cal
The sensor board was not calibrated at
the factory.
Sensor Out of Range
The pH measurement is invalid.
Sensor Negative Reading
The chlorine reading is less than -0.5
ppm.
Sensor RTD Sense Open
RTD sensor line is broken or not
connected.
Sensor Temperature High
Temperature is greater than 155 °C
(311 °F).
Sensor Temperature High or Low
Sensor Temperature Low
Temperature is less than -20 °C (-4 °F).
Sensor Temperature High or Low
Broken Glass Disabled
Advisory only (applies to pH sensor
only).
9.4.1
Sensor No Solution Gnd
This message implies that the pH sensor is miswired. Check sensor wiring.
9.4.2
Sensor Need Factory Cal
The sensor board was improperly calibrated at the factory. Call the factory for assistance.
Reference Manual
Troubleshoot
00809-0100-3412
May 2019
Rosemount FCL 1056
91
Summary of Contents for Rosemount FCL
Page 4: ...4 ...
Page 18: ...Install Reference Manual May 2019 00809 0100 3412 18 Emerson com Rosemount ...
Page 26: ...Wire Reference Manual May 2019 00809 0100 3412 26 Emerson com Rosemount ...
Page 27: ...Reference Manual Wire 00809 0100 3412 May 2019 Rosemount FCL 1056 27 ...
Page 28: ...Wire Reference Manual May 2019 00809 0100 3412 28 Emerson com Rosemount ...
Page 76: ...Digital communications Reference Manual May 2019 00809 0100 3412 76 Emerson com Rosemount ...
Page 107: ...Reference Manual 00809 0100 3412 May 2019 Rosemount FCL 1056 107 ...