1. ±0.4 °C accuracy is not acceptable.
2. The temperature measurement is suspected of being in error. Calibrate
temperature by making the transmitter reading match the temperature measured
with a standard thermometer.
Procedure
1. Remove the sensor from the flow cell. Place it in an insulated container of water
along with a calibrated thermometer. Submerge at least the bottom two inches of
the sensor.
2. Allow the sensor to reach thermal equilibrium.
The time constant for the sensor is about five minutes, so it may take as long as
thirty minutes for equilibration.
3. Press
MENU
.
The main
Menu
screen appears. The cursor is on Calibrate.
4. Press
ENTER
.
5. Choose the sensor you wish to calibrate.
Sensor 1 is the chlorine sensor. Sensor 2 (if present) is the pH sensor.
6. Choose Temperature.
7. Change the display to match the temperature read from the calibrated
thermometer. Press
ENTER
.
If the present temperature is more than 5 °C different from the value entered, an
error message appears.
8. To force the transmitter to accept the calibration, choose Yes. To repeat the
calibration, choose No.
For troubleshooting assistance, see
Troubleshooting when no error message is
9. To return to the main display, press
MENU
and then
EXIT
.
6.3
Calibration - free chlorine
6.3.1
Purpose
shows, a free chlorine sensor generates a current directly proportional to the
concentration of free chlorine in the sample. Calibrating the sensor requires exposing it to
a solution containing no free chlorine (zero standard) and to a solution containing a known
amount of free chlorine (full-scale standard).
Calibrate
Reference Manual
May 2019
00809-0100-3412
64
Emerson.com/Rosemount
Summary of Contents for Rosemount FCL
Page 4: ...4 ...
Page 18: ...Install Reference Manual May 2019 00809 0100 3412 18 Emerson com Rosemount ...
Page 26: ...Wire Reference Manual May 2019 00809 0100 3412 26 Emerson com Rosemount ...
Page 27: ...Reference Manual Wire 00809 0100 3412 May 2019 Rosemount FCL 1056 27 ...
Page 28: ...Wire Reference Manual May 2019 00809 0100 3412 28 Emerson com Rosemount ...
Page 76: ...Digital communications Reference Manual May 2019 00809 0100 3412 76 Emerson com Rosemount ...
Page 107: ...Reference Manual 00809 0100 3412 May 2019 Rosemount FCL 1056 107 ...