2. To assign values to the analog outputs, you must open the
Program
sub-menu. Use
Down
to move the cursor to Program. Press
ENTER
.
The
Program
menu appears. There are between five and seven items in the
Program
menu. Diagnostic Setup appears only if you have the Rosemount
™
FCL-02
with pH sensor. The screen displays four items at a time. The downward pointing
arrow on the right of the screen shows there are more items available in the menu.
3. To view the other items, use
Down
to scroll to the last item shown and continue
scrolling down. When you have reached the bottom, the arrow will point up. Move
the cursor back to Outputs and press
ENTER
.
The
Outputs
screen appears. The cursor is on Range. Output range is used to assign
values to the low and high current outputs.
4. Press
ENTER
.
The
Output Range
screen appears. The screen shows the present values assigned
to output 1 (O1) and output 2 (O2). The screen also shows which sensors the
outputs are assigned to. S1 is sensor 1.. The assignments shown are the defaults for
the Rosemount FCL-01. Outputs are freely assignable under the
Configure
menu.
5. For practice, change the 20 mA settings for output 1 to
8.5 ppm
.
a) Move the cursor to the O1 S1 20 mA: 10.00 line and press
ENTER
.
The screen below appears.
Reference Manual
Display and operation
00809-0100-3412
May 2019
Rosemount FCL 1056
33
Summary of Contents for Rosemount FCL
Page 4: ...4 ...
Page 18: ...Install Reference Manual May 2019 00809 0100 3412 18 Emerson com Rosemount ...
Page 26: ...Wire Reference Manual May 2019 00809 0100 3412 26 Emerson com Rosemount ...
Page 27: ...Reference Manual Wire 00809 0100 3412 May 2019 Rosemount FCL 1056 27 ...
Page 28: ...Wire Reference Manual May 2019 00809 0100 3412 28 Emerson com Rosemount ...
Page 76: ...Digital communications Reference Manual May 2019 00809 0100 3412 76 Emerson com Rosemount ...
Page 107: ...Reference Manual 00809 0100 3412 May 2019 Rosemount FCL 1056 107 ...