8.4.2
Cleaning the flow controller
The flow controller can be taken apart completely for cleaning.
Procedure
1. Use a strong flow of water to flush out the tubing.
Use a pipe cleaner or small bottlebrush to remove more adherent deposits.
2. To prevent leaks, apply a thin layer of silicone grease (or equivalent) to the two O-
rings as the base of the overflow chamber and to the O-ring sealing the central
overflow tube to the base.
8.4.3
Other maintenance
show the replacement parts for the flow controller assembly used
in the Rosemount FCL-01.
and
show replacement parts for the flow
controller assembly used in the Rosemount FCL-02.
Table 8-2: Rosemount FCL-01 Constant Head Flow Controller Assembly Replacement Parts
Location in
PN
Description
1
24039-00
Flow cell for chlorine sensor with bubble shedding nozzle
2
24040-00
O-ring kit, two 2-222 and one 2-024 silicone O-rings with lubricant
3
33812-00
Dust cap for constant head flow controller
4
9322032
Elbow, 1/4 in. FNPT x 1/4 in. OD tubing
5
9350029
Check valve, 1/4 in. FNPT
6
33823-00
Outside tube for constant head device
Maintenance
Reference Manual
May 2019
00809-0100-3412
82
Emerson.com/Rosemount
Summary of Contents for Rosemount FCL
Page 4: ...4 ...
Page 18: ...Install Reference Manual May 2019 00809 0100 3412 18 Emerson com Rosemount ...
Page 26: ...Wire Reference Manual May 2019 00809 0100 3412 26 Emerson com Rosemount ...
Page 27: ...Reference Manual Wire 00809 0100 3412 May 2019 Rosemount FCL 1056 27 ...
Page 28: ...Wire Reference Manual May 2019 00809 0100 3412 28 Emerson com Rosemount ...
Page 76: ...Digital communications Reference Manual May 2019 00809 0100 3412 76 Emerson com Rosemount ...
Page 107: ...Reference Manual 00809 0100 3412 May 2019 Rosemount FCL 1056 107 ...