Figure 9-5: Simulating Resistance Temperature Device Inputs
A. Resistance temperature device return
B. Resistance temperature device sense
C. Resistance temperature device in
D. Resistance temperature device shield
To check the accuracy of the temperature measurement, set the resistor simulating the
resistance temperature device to the values indicated in the table and note the
temperature readings. The measured temperature might not agree with the value in the
table. During sensor calibration, an offset might have been applied to make the measured
temperature agree with a standard thermometer. The offset is also applied to the
simulated resistance. The transmitter is measuring temperature correctly if the difference
between measured temperatures equals the difference between the values in the table to
within ±0.1 °C.
For example, start with a simulated resistance of 103.9 Ω, which corresponds to 10.0 °C.
Assume the offset from the sensor calibration was -0.3 Ω. Because of the offset, the
transmitter calculates temperature using 103.6 Ω. The result is 9.2 °C. Now change the
resistance to 107.8 Ω, which corresponds to 20.0 °C. The transmitter uses 107.5 Ω to
calculate the temperature, so the display reads 19.2 °C. Because the difference between
the displayed temperatures (10.0 °C) is the same as the difference between the simulated
temperatures, the transmitter is working correctly.
Temp. (°C)
Pt 100 (Ω)
0
100.0
10
103.9
20
107.8
25
109.7
30
111.7
40
115.5
50
119.4
Reference Manual
Troubleshoot
00809-0100-3412
May 2019
Rosemount FCL 1056
105
Summary of Contents for Rosemount FCL
Page 4: ...4 ...
Page 18: ...Install Reference Manual May 2019 00809 0100 3412 18 Emerson com Rosemount ...
Page 26: ...Wire Reference Manual May 2019 00809 0100 3412 26 Emerson com Rosemount ...
Page 27: ...Reference Manual Wire 00809 0100 3412 May 2019 Rosemount FCL 1056 27 ...
Page 28: ...Wire Reference Manual May 2019 00809 0100 3412 28 Emerson com Rosemount ...
Page 76: ...Digital communications Reference Manual May 2019 00809 0100 3412 76 Emerson com Rosemount ...
Page 107: ...Reference Manual 00809 0100 3412 May 2019 Rosemount FCL 1056 107 ...