System Alarms
(continued)
Table 7-3:
Alarm label
Help messages and recommended actions
SB Mismatch
The sensor type does not match, indicating wrong sensor
board is connected.
SB Version
The sensor board software version does not match the ex-
pected version, indicating the sensor board software needs
to be upgraded. Place the jumper on J1 of the sensor board
and cycle the power to start the sensor board software up-
grade.
SB Bootload Mode
The Bootload Mode jumper at J1 of the sensor board is on
at power up. This mode should only be used to upgrade
the sensor board software. Remove the jumper from J1 of
the sensor board and cycle the power to clear the alarm.
Probe Alarms
Table 7-4:
Alarm label
Help messages and recommended actions
O
2
T/C Reversed
The sensor heater thermocouple voltage is reading a nega-
tive voltage, indicating the thermocouple wire connec-
tions may be reversed. Check wiring.
NV Memory Fail
Checksum error was detected in the nonvolatile memory
configuration data when the unit was turned on. Default
values have been loaded. Check to see that your configura-
tions have not been changed. Cycle the power to clear
alarm.
Board Temp High
The electronics board tempreature reading is above the
threshold. The board may not function correctly.
Factory Mode
The Factory Mode switch is set to
On
at power up. This
mode should only be uses for factory setup. Turn off the
Factory Mode switch and cycle the power to clear alarm.
Program Integrity
Software checksum error detected when the unit was
turned on. Please replace sensor board.
O
2
T/C Shorted
This diagnostic is only intended to detect a copper short of
the thermocouple connections at the electronics connec-
tor. The cell heater T/C voltage is reading close to zero, in-
dicating the thermocouple wires may be shorted.
O
2
Sensor Open
The cell impedance voltage is reading out of normal range,
indicating the sensor wires may be disconnected or the
sensor junction may be open. Check wiring.
O
2
T/C Open
The sensor heater thermocouple voltage is out of range,
indicating the T/C wires may be disconnected, or the T/C
junction may be open. Check wiring.
O
2
Temp Low
The sensor heater temperature is below the low tempera-
ture threshold.
Parameters
Reference Manual
35
Summary of Contents for Rosemount CX1100
Page 1: ...Reference Manual 00809 0100 4110 Rev AA June 2018 Rosemount CX1100 In Situ Oxygen Transmitter ...
Page 4: ......
Page 6: ...Contents ii Rosemount CX1100 ...
Page 18: ...Wall Surface and Pipe Mount Figure 2 4 Install 12 Rosemount CX1100 ...
Page 19: ...Panel Mount Figure 2 5 Install Reference Manual 13 ...
Page 20: ...Install 14 Rosemount CX1100 ...
Page 26: ...Wire 20 Rosemount CX1100 ...
Page 28: ...Startup 22 Rosemount CX1100 ...
Page 32: ...Calibration and maintenance 26 Rosemount CX1100 ...
Page 36: ...Rosemount CX1100 In Situ Oxygen Transmitter menu trees 30 Rosemount CX1100 ...
Page 45: ...Rosemount CX1100 product certifications Reference Manual 39 ...
Page 46: ...Rosemount CX1100 product certifications 40 Rosemount CX1100 ...
Page 47: ...Rosemount CX1100 product certifications Reference Manual 41 ...