Probe Parameters
(continued)
Table 7-2:
Parameter label
Description
O
2
LRV
Primary variable, O
2
%, lower range value. (
0.0%
only)
O
2
URV
Primary variable, O
2
% upper range value (
1.0%
-
50.0%
)
AO Range
Analog ouput polarity
Alarm Level
O
2
alarm level
Relay 1
Alarm Relay 1 mode
Relay 2
Alarm Relay 2 mode
AO Track
Analog output track O
2
sensor measurement
during a calibration
Cal Gas 1
Test Gas 1 value. This is the actual value of the
gas being applied during the Test Gas 1 phase
of a calibration. (
0.05
-
50.0%
)
Cal Gas 2
Test Gas 2 value. this is the actual value of the
gas being applied during the Test Gas 2 phase
of calibration. (
0.05
-
50.0%
)
Gas Time
Test gas application time. This is the length of
time test gases are applied to the O
2
probe dur-
ing low or high test gas phase of a calibration.
(
60
-
1200 sec
)
Purge Time
Test gas purge time. This is the length of time
before the output will be returned to the proc-
ess reading after a calibration. (
60
-
1200 sec
)
Imp Check Hold
Time to hold analog output during impedance
check. (
0
-
60 sec
)
(Factory
mode only)
Remote CJC
Use local or remote CJC measurement for cor-
rection. (
No
/
Yes
) (
Factory
mode only)
Raw Counts | Cell Voltage
O
2
sensor voltage raw counts. (
Factory
mode
only)
Raw Counts | Cell Imp
Cell impedance/sensor resistance raw counts.
(
Factory
mode only)
Raw Counts | T/C Voltage
O
2
T/C raw counts. (
Factory
mode only)
Raw Counts | Heater Voltage
Heater voltage raw counts. (
Factory
mode only)
Cal State
This represents the state of the calibration cycle
it is in.
Time Remain
Time remaining in the present calibration cycle
state.
Slope
Current calibration slope. This is the slope value
that was calculated as a result of the last suc-
cessful calibration. (
34.5
to
57.5 mV/Dec
)
Parameters
Reference Manual
33
Summary of Contents for Rosemount CX1100
Page 1: ...Reference Manual 00809 0100 4110 Rev AA June 2018 Rosemount CX1100 In Situ Oxygen Transmitter ...
Page 4: ......
Page 6: ...Contents ii Rosemount CX1100 ...
Page 18: ...Wall Surface and Pipe Mount Figure 2 4 Install 12 Rosemount CX1100 ...
Page 19: ...Panel Mount Figure 2 5 Install Reference Manual 13 ...
Page 20: ...Install 14 Rosemount CX1100 ...
Page 26: ...Wire 20 Rosemount CX1100 ...
Page 28: ...Startup 22 Rosemount CX1100 ...
Page 32: ...Calibration and maintenance 26 Rosemount CX1100 ...
Page 36: ...Rosemount CX1100 In Situ Oxygen Transmitter menu trees 30 Rosemount CX1100 ...
Page 45: ...Rosemount CX1100 product certifications Reference Manual 39 ...
Page 46: ...Rosemount CX1100 product certifications 40 Rosemount CX1100 ...
Page 47: ...Rosemount CX1100 product certifications Reference Manual 41 ...