2
Install
WARNING!
ELECTRIC SHOCK
Install all protective equipment covers and safety ground leads after installation. Failure to
install covers and ground leads could result in serious injury or death.
CAUTION!
EQUIPMENT DAMAGE
Do not allow the temperature of the Rosemount
™
CX1100 Probe electronics to exceed 194 °F
(90 °C) or damage to the unit may result.
2.1
Probe installation
A
weld plate
for welding to the flue gas duct can be supplied for new installations.
1.
If using the standard square weld plate (
Figure 2-2
) or an optional flange mounting
plate, weld or bolt the plate onto the duct.
The through hole diameter in the stack or duct wall and refractory material must be
at least 2.5 in. (63.5 mm).
2.
Insert probe through the opening in the mounting flange and bolt through the
probe gasket and flange.
For horizontal installations, the breather port must be oriented such that it is facing
downward.
Install
Reference Manual
7
Summary of Contents for Rosemount CX1100
Page 1: ...Reference Manual 00809 0100 4110 Rev AA June 2018 Rosemount CX1100 In Situ Oxygen Transmitter ...
Page 4: ......
Page 6: ...Contents ii Rosemount CX1100 ...
Page 18: ...Wall Surface and Pipe Mount Figure 2 4 Install 12 Rosemount CX1100 ...
Page 19: ...Panel Mount Figure 2 5 Install Reference Manual 13 ...
Page 20: ...Install 14 Rosemount CX1100 ...
Page 26: ...Wire 20 Rosemount CX1100 ...
Page 28: ...Startup 22 Rosemount CX1100 ...
Page 32: ...Calibration and maintenance 26 Rosemount CX1100 ...
Page 36: ...Rosemount CX1100 In Situ Oxygen Transmitter menu trees 30 Rosemount CX1100 ...
Page 45: ...Rosemount CX1100 product certifications Reference Manual 39 ...
Page 46: ...Rosemount CX1100 product certifications 40 Rosemount CX1100 ...
Page 47: ...Rosemount CX1100 product certifications Reference Manual 41 ...