The Rosemount CX1100 Remote Transmitter is available in a panel mounting or wall/pipe
mounting configuration. Refer to
Figure 2-4
or
Figure 2-5
for the panel, wall, or pipe
mounting details. You need a wrench and bolts to mount the transmitter.
Procedure
1.
Ensure all components are available to install the Rosemount CX1100 Remote
Transmitter.
2.
Select a mounting location near or removed from the Rosemount CX1100 Probe.
Consider the temperature limitations of the Rosemount CX1100 Remote
Transmitter. Refer to product specifications when selecting the mounting location.
3.
Mount at a height convenient for viewing and operating the interface.
Emerson recommends approximately 5 ft. (1.5 m).
4.
The keypad window on the Rosemount CX1100 Remote Transmitter may have an
exterior protective membrane. Remove the protective membrane prior to use of the
Rosemount CX1100 Remote transmitter.
Failure to remove the protective membrane may cause the display to appear
distorted. The membrane may become difficult or impossible to remove after
extended use at elevated temperatures.
Install
Reference Manual
11
Summary of Contents for Rosemount CX1100
Page 1: ...Reference Manual 00809 0100 4110 Rev AA June 2018 Rosemount CX1100 In Situ Oxygen Transmitter ...
Page 4: ......
Page 6: ...Contents ii Rosemount CX1100 ...
Page 18: ...Wall Surface and Pipe Mount Figure 2 4 Install 12 Rosemount CX1100 ...
Page 19: ...Panel Mount Figure 2 5 Install Reference Manual 13 ...
Page 20: ...Install 14 Rosemount CX1100 ...
Page 26: ...Wire 20 Rosemount CX1100 ...
Page 28: ...Startup 22 Rosemount CX1100 ...
Page 32: ...Calibration and maintenance 26 Rosemount CX1100 ...
Page 36: ...Rosemount CX1100 In Situ Oxygen Transmitter menu trees 30 Rosemount CX1100 ...
Page 45: ...Rosemount CX1100 product certifications Reference Manual 39 ...
Page 46: ...Rosemount CX1100 product certifications 40 Rosemount CX1100 ...
Page 47: ...Rosemount CX1100 product certifications Reference Manual 41 ...