calibration values do not meet the accepted performance criteria, the existing calibration
values remain in effect, and Emerson recommends replacing the Rosemount CX1100
Probe.
CAUTION!
EQUIPMENT DAMAGE
Make sure that the calibration gas cap is replaced tightly after calibration is complete. Many
combustion processes operate at a slight negative pressure (draft pressure) and can draw
ambient air down the cal gas lines and into the sensing cell, causing a false elevated oxygen
reading. The same phenomenon is possible if the calibration gas hoses become degraded or
loose.