Instruction Manual
IM-105-4000, Rev 1.3
October 2008
6-5
OPM 4000
1. Remove the transceiver and retro reflector from the hinge pins.
2. Remove the service module and install the system on test stands at the
correct flange-to-flange distance plus 11 in. (279,4 mm).
The additional 11 in. (279,4 mm) compensates for air plenum spacing,
as the air plenums are not used for the off stack zero calibration. If test
stands are not available, an alternate method is to use 3 in. pipe flanges
with air plenums installed. Contact the factory for air plenum availability.
Set the spacing between the 3 in. flanges exactly the same as the
flange-to-flange measurement.
3. Clean the transceiver and retro reflector windows.
4. Connect the control unit with the control to service module test cable kit
and apply power to the system.
5. The retro reflector must be level.
6. Follow instruction for "Clear on Stack Zero and Span Calibration".
Summary of Contents for Rosemount Analytical OPM 4000
Page 2: ......
Page 6: ......
Page 10: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 TOC 4 ...
Page 14: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 iv ...
Page 42: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 2 20 ...
Page 52: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 4 6 ...
Page 58: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 5 6 ...
Page 64: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 6 6 ...
Page 100: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 A 24 ...
Page 102: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 B 2 ...
Page 104: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 C 2 ...
Page 114: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 D 10 ...
Page 116: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 Index 2 OPM 4000 ...