Instruction Manual
IM-105-4000, Rev 1.3
October 2008
2-11
OPM 4000
Mounting the
Control Unit
the OPM 4000 control unit is designed for rack mounting in a customer
supplied control panel. Install the OPM 4000 control unit in the designated
control panel according to the following procedure.
NOTE
The control unit (Figure 2-8) should be mounted at eye level for best viewing
of the display.
1. Make a 3.13 x 7.5 in. (80 x 191 mm) cut-out in the control panel where
the control unit will be mounted (Figure 2-9).
2. Check that gasket seals are snug against each side of the flange, as
3. Slide the controller into the cut-out.
4. From the rear of the control unit, insert the panel mounting hardware in
the slots provided on the top and bottom of the control unit
(Figure 2-10).
5. Secure the control unit by tightening the screws with the 1/8 in. hex
socket wrench until the control unit is securely held in place
(Figure 2-10).
Figure 2-8. Control Unit
38450007
Summary of Contents for Rosemount Analytical OPM 4000
Page 2: ......
Page 6: ......
Page 10: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 TOC 4 ...
Page 14: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 iv ...
Page 42: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 2 20 ...
Page 52: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 4 6 ...
Page 58: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 5 6 ...
Page 64: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 6 6 ...
Page 100: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 A 24 ...
Page 102: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 B 2 ...
Page 104: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 C 2 ...
Page 114: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 D 10 ...
Page 116: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 Index 2 OPM 4000 ...