Instruction Manual
IM-105-4000, Rev 1.3
October 2008
OPM 4000
2-10
Attach Blowers
Connect an air blower to the air blower port on the transceiver using the air
hoses and clamps. Connect the other air blower to the air blower port on the
retro reflector using the air hoses and clamps. See Figure 2-6 for the air
blower port location.
Mounting the
Transceiver and Retro
Reflector
Install the transceiver and retro reflector according to the following procedure.
This installation procedure applies to both the transceiver and the retro
reflector. Installation dimensions and clearances are shown in Figure 2-1.
1. Power up the air-purge blowers to prevent stack particulate from
accumulating in the nipple and air-purge housing.
2. Attach the transceiver and retro reflector to the air plenum assembly by
placing them on the hinge pins (Figure 2-7).
3. Close transceiver and retro reflector and secure in place with the two
hold-down latches.
Figure 2-7. Upper and Lower
Hinge Pins
If installed location has a positive pressure the air-purge system must be used continuously
during installation to prevent process gases from contaminating optical surfaces or over
heating instrument electronics. If the system is shut off for more than momentary
interruptions, the instrument may be damaged. Failure to provide continuous air-purge may
void the warranty.
Upper Hinge Pin
Lower Hinge Pin
38450004
Summary of Contents for Rosemount Analytical OPM 4000
Page 2: ......
Page 6: ......
Page 10: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 TOC 4 ...
Page 14: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 iv ...
Page 42: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 2 20 ...
Page 52: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 4 6 ...
Page 58: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 5 6 ...
Page 64: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 6 6 ...
Page 100: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 A 24 ...
Page 102: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 B 2 ...
Page 104: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 C 2 ...
Page 114: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 D 10 ...
Page 116: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 Index 2 OPM 4000 ...