Instruction Manual
760000-C
November 2002
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-9
Model NGA2000 CLD
MLT/CH1
Display
Status…
Main…
Channel
BasicCal
0.00 Range: 2 5.00
2.50
% CO2
Temperature: 37.0 C
0.0
100.0
Maintenance-Request No
Any-alarms: No
Operation: Ready
Select
Status…
Tags Off
LCDReset
7.50
2.50
95.00
CLD
ppm NO
MLT/CH1
% CO2
MLT/CH2
ppm CO
0.00 [1]
10.00
0.00
0 [2] 5.00
0.00
[2] 250.00
>
CLD
Sample flow: 1300 ml/min 200 1500
Sample press: 4.0 hPa 50.0 490.0
Ozonator: OFF – PRESS. SW.
Converter temp: 30.9 C 150.0 500.0
Display
Status…
Main…
Channel
BasicCal
0.00 Range: 1 10.00
7.50
ppm NO
c. Multi
Component
Display
If other analyzer modules are connected
to the Platform, it is possible to display up
to five using the following steps to change
from the single component display to the
multi component display as follows:
From the single channel display (Meas-
ure) press Display (F1) to change to the
Multi Component Display of all other in-
stalled Analyzer Modules.
Changing to the multi component display
can be done from each single component
display.
Each bargraph shows the start and end of
the range for the respective channel. The
number in parentheses indicates the
number of the selected range for that
channel. (F.S. = full scale)
(Display may look different depending on
installed analyzers.)
Use the F3 softkey (Tags Off) to turn the
analyzer tags on or off.
To select a single channel display in the
multi channel display, enable the select
symbol (>) by pressing the F1 softkey or
the
↓
softkey.
Then use the
↓
or
↑
softkey to select the
line for the desired channel. When the
desired channel is marked, select it for
single component display by pressing the
F1 softkey.