Instruction Manual
760000-C
November 2002
3-8 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 CLD
CLD
7.50 ppm
HOME
ESCAPE
MORE
INFO
-- Current measurement parameters --
Response time: 3.0 s
Bypass flow: 1300 ml/min.
Capillary pressure: 4.0 hPa
Detector pressure: 38.8 c
CLD
7.50 ppm
HOME
ESCAPE
MORE
INFO
-- Current measurement parameters --
Analyzer gas measured: NO
Measurement range number: 1
Range change control: Local
Linearization mode: Disabled
Analyzer operational state: STANDBY
Analyzer alarm state: NORMAL
Alarms reported: FAILURE
Current total variable updates per second: 24
CLD
Sample flow: 1300 ml/min 200 1500
Sample press: 4.0 hPa 50.0 490.0
Ozonator: OFF – PRESS. SW.
Converter temp: 30.9 C 150.0 500.0
Display
Status…
Main…
Channel
BasicCal
0.00 Range: 1 10.00
7.50
ppm NO
MLT/CH1/R2
Display
Status…
Main…
Channel
BasicCal
Temperature: 37.0 C 0.0 100.0
Maintenance Requests: No
Any alarms: No
Operation: Ready
0.00 Range: 2 5.00
2.50
% CO2
CLD
Sample flow: 1300 ml/min 200 1500
Sample press: 4.0 hPa 50.0 490.0
Ozonator: OFF – PRESS. SW.
Converter temp: 30.9 C 150.0 500.0
Display
Status…
Main…
Channel
BasicCal
0.00 Range: 1 10.00
7.50
ppm NO
3-4 BASIC CONTROLS AND STATUS
The following sections describe the basic con-
trol of the analyzer and the viewing of channel
parameters. Examples of stepping through the
menus are shown so that the user can be-
come familiar with the operation, keeping in
mind that displays and menu choices may be
different depending on actual analyzer con-
figuration and any customization of the
menus.
a. Analyzer Channel Status
The analyzer status is displayed in the
Current measurement parameters dis-
play.
To access the Current measurement pa-
rameters display:
Main Menu [Measure (F1)]
↓
Measurement Mode [STATUS(F2)]
↓
Current measurement parameters
From the Current measurement parame-
ters menu an additional sub menu is
available with additional parameters.
Press MORE (F3) to access the next
menu.
Press ESCAPE (F2) or MORE (F4) to re-
turn to the first menu or HOME (F1) to re-
turn to the Main menu display.
b. Single
Component
Display
The Measure menu that displays after
startup is the Single Component display
of the analyzer. If other analyzer modules
are connected to the Platform, it is possi-
ble to display them using the following
steps to change the channel of the single
component display:
From the Measure Mode display, press
Channel (F4) to change to the Single
Component Display of any other installed
Analyzer Modules.
Example: Changing from CLD (channel 1)
to CO
2
(channel 2). Continue pressing
Channel (F4) to display the desired chan-
nel depending on the installed analyzer
configuration, ultimately returning to the
first channel.