reactive as long as the battery and the power module are maintained. Care should be
taken to prevent thermal, electrical, or mechanical damage. Contacts should be protected
to prevent premature discharge.
CAUTION
Use caution when handling the power module, it may be damaged if dropped from
heights exceeding 6.10 m (20 ft).
3.2.5
Environmental
Verify the operating atmosphere of the transmitter is consistent with the appropriate
hazardous locations certification.
Temperature effects
The transmitter will operate within specifications for ambient temperatures between –40
and 85 °C (–40 and 185 °F).
Heat from the process is transferred to the transmitter housing. If the process
temperature is high, the ambient temperature will need to be lower to account for heat
transferred to the transmitter housing. See
for ambient temperature derating.
Refer to
for High Temperature option (HT01) for ambient temperature derating.
Table 3-1: Ambient Temperature Derating
Process temperature
Max ambient
°C
°F
°C
°F
260
500
41
106
240
464
45
113
220
428
49
120
200
392
53
127
180
356
57
135
160
320
61
142
140
284
64
147
120
248
58
136
100
212
72
162
85
185
75
167
Table 3-2: Ambient Temperature Derating - HT01
Process temperature
Max ambient
°C
°F
°C
°F
550
1022
41
106
520
968
45
113
490
914
47
117
Installation
Reference Manual
June 2022
00809-0100-4708
16