Recommended actions
1. Verify environmental temperature is within the transmitter’s range.
2. Remote mount the transmitter away from process and environmental
conditions.
3. Reset the device.
4. If the condition persists, replace the electronics.
Terminal temperature has exceeded limits
Description
The terminal temperature has exceed the transmitter's maximum range.
Recommended actions
1. Verify environmental temperature is within the transmitter’s range.
2. Remote mount the transmitter away from process and environmental
conditions.
3. Reset the device.
4. If the condition persists, replace the electronics.
Supply voltage low
Description
The supply voltage is low and may soon affect broadcast updates.
Recommended actions
Replace the power module.
Database memory warning
Description
The device has failed to write to the database memory. Any data written during this time
may have been lost.
Recommended actions
1. Reset the device.
2. Reconfirm all configuration items in the device.
3. If logging dynamic data not needed, this advisory can be safely ignored.
4. If the condition persists, replace the electronics.
Invalid Configuration
Description
The device has detected a configuration error based on a change to the device.
Reference Manual
Troubleshooting
00809-0200-4648
May 2020
Rosemount 648 Wireless
61
Summary of Contents for Rosemount 648
Page 4: ...4 ...
Page 8: ...Introduction Reference Manual May 2020 00809 0200 4648 8 Rosemount 648 Wireless ...
Page 32: ...Configuration Reference Manual May 2020 00809 0200 4648 32 Rosemount 648 Wireless ...
Page 48: ...Commissioning Reference Manual May 2020 00809 0200 4648 48 Rosemount 648 Wireless ...
Page 58: ...Operation and maintenance Reference Manual May 2020 00809 0200 4648 58 Rosemount 648 Wireless ...
Page 66: ...Troubleshooting Reference Manual May 2020 00809 0200 4648 66 Rosemount 648 Wireless ...
Page 68: ...Reference data Reference Manual May 2020 00809 0200 4648 68 Rosemount 648 Wireless ...