Instruction Manual
Section 9: Alarm Relay Functions
LIQ-MAN-56
April 2017
Alarm Relay Functions
67
Section 9.0 Alarm Relay Functions
9.1
General
An alarm is a relay that closes a set of contact points (a switch) inside the analyzer. In doing
so, the relay closes an electrical circuit and turns on a device wired to the contacts. The 56
Advanced Analyzer has four alarm relays and seven relay control functions. The relays are
turned on and off by the analyzer based on the control points, setpoints or control
parameters that you program into the analyzer through the keypad. See Section 9.2
through 9.7 to program the alarm relay functions. Each relay functions section includes a
description, a figure detailing its operation, a step-by-step setup procedure, and a table or
default and programmable limit settings.
The 56 has the following relay control functions:
9.2
High/Low Concentration Alarm
9.2.1
Description
High/Low concentration alarms are setpoint alarms with adjustable deadband. These
operate as simple on/off alarms used for applications requiring discrete on/off control of
pumps and valves. Typical applications include demineralizer bed regeneration and
blowdown in boilers and cooling towers. Any active device variable in the 56 analyzer can be
programmed as a high/low concentration alarm including the primary or secondary
variables, temperature, raw values and diagnostics.
A schematic of the high/low concentration (setpoint) alarm operation is shown below.
Relay Control Functions
Common applications
Section
High/Low Concentration Alarm
measurement setpoint control
Sec. 9.2
Delay Timer
chemical mixing and neutralization
Sec. 9.3
Bleed and Feed
blowdown and chemical addition
Sec. 9.4
Totalizer Relay Activation
chemical dosing in reactors
Sec. 9.5
Interval Timer
periodic probe cleaning
Sec. 9.6
Date and Time Activation
seawater-cooled condensers
Sec. 9.7
Figure 9-1. High/Low Concentration Alarm operation
high alarm
t
n
i
o
p
t
e
s
high alarm logic
1.5
1.3
deadband (0.2)
time
PV
alarm activates
alarm deactivates
Table 9-1. Alarm Relay Functions
Summary of Contents for Rosemount 56
Page 1: ...Instruction Manual LIQ MAN 56 Rev D April 2017 Rosemount 56 Advanced Dual Input Analyzer ...
Page 2: ......
Page 10: ...iv Table of Contents Instruction Manual April 2017 LIQ MAN 56 ...
Page 30: ...Section 2 Installation Instruction Manual April 2017 LIQ MAN 56 20 Installation ...
Page 40: ...30 Wiring Section 3 Wiring Instruction Manual April 2017 LIQ MAN 56 ...
Page 64: ...Section 6 Programming Measurements Instruction Manual April 2017 LIQ MAN 56 54 ...
Page 76: ...Section 8 Time Proportional Control Instruction Manual April 2017 LIQ MAN 56 66 ...
Page 124: ...Section 13 Maintenance Instruction Manual April 2017 LIQ MAN 56 114 ...
Page 126: ...Section 14 Return of Material Instruction Manual April 2017 LIQ MAN 56 116 ...