Profibus Communications
99
Instruction Manual
Section 12: Profibus Communications
LIQ-MAN-56
April 2017
Diagnostic Information
The 56 provides diagnostic information on Profibus communications, which is accessed by
pressing the INFO key and then selecting “Profibus information” to launch the window below:
•
Address –
displays the address being used by the 56 on the network
•
Baud rate –
the baud rate of the 56 and the network
•
Master detected –
If the master is not detected, the host and analyzer are not commu-
nicating. Check wiring connections, make sure the addresses match, and try changing
the baud rate.
•
POST, Lpf output, Rx character count –
POST is the power-on self test. POST and the
remaining parameters on this window are used as factory diagnostics only.
Pressing the NEXT button brings up the final diagnostic parameter, the software version used
by the Profibus board:
Summary of Contents for Rosemount 56
Page 1: ...Instruction Manual LIQ MAN 56 Rev D April 2017 Rosemount 56 Advanced Dual Input Analyzer ...
Page 2: ......
Page 10: ...iv Table of Contents Instruction Manual April 2017 LIQ MAN 56 ...
Page 30: ...Section 2 Installation Instruction Manual April 2017 LIQ MAN 56 20 Installation ...
Page 40: ...30 Wiring Section 3 Wiring Instruction Manual April 2017 LIQ MAN 56 ...
Page 64: ...Section 6 Programming Measurements Instruction Manual April 2017 LIQ MAN 56 54 ...
Page 76: ...Section 8 Time Proportional Control Instruction Manual April 2017 LIQ MAN 56 66 ...
Page 124: ...Section 13 Maintenance Instruction Manual April 2017 LIQ MAN 56 114 ...
Page 126: ...Section 14 Return of Material Instruction Manual April 2017 LIQ MAN 56 116 ...