pH-Independent Free Chlorine
The 56 is compatible with the 498CL-01 pH-independent free chlorine sensor. The 498CL-01
sensor is intended for the continuous determination of free chlorine (hypochlorous acid plus
hypochlorite ion) in water. The primary application is measuring chlorine in drinking water. The
sensor requires no acid pre-treatment, nor is an auxiliary pH sensor required for pH correction.
The 56 fully compensates free chlorine readings for changes in membrane permeability caused
by temperature. For more information concerning the use and operation of the
amperometric chlorine sensors, refer to the product data sheets.
Performance Specifications - Analyzer
Resolution:
0.001 ppm or 0.01 ppm – selectable
Input Range:
0 nA – 100 µA
pH independent
Temperature compensation:
Automatic (via RTD) or manual (0-50 °C).
Input filter:
Time constant 1 - 999 sec, default 5 sec.
Response time:
6 seconds to 95% of final reading
Recommended Sensors
Rosemount 498CL-01 pH independent free chlorine sensor
Chlorine sensors with Variopol
connection and cable connection
498CL-01
Description and Specifications
11
Instruction Manual
Section 1: Description and Specifications
LIQ-MAN-56
April 2017
Summary of Contents for Rosemount 56
Page 1: ...Instruction Manual LIQ MAN 56 Rev D April 2017 Rosemount 56 Advanced Dual Input Analyzer ...
Page 2: ......
Page 10: ...iv Table of Contents Instruction Manual April 2017 LIQ MAN 56 ...
Page 30: ...Section 2 Installation Instruction Manual April 2017 LIQ MAN 56 20 Installation ...
Page 40: ...30 Wiring Section 3 Wiring Instruction Manual April 2017 LIQ MAN 56 ...
Page 64: ...Section 6 Programming Measurements Instruction Manual April 2017 LIQ MAN 56 54 ...
Page 76: ...Section 8 Time Proportional Control Instruction Manual April 2017 LIQ MAN 56 66 ...
Page 124: ...Section 13 Maintenance Instruction Manual April 2017 LIQ MAN 56 114 ...
Page 126: ...Section 14 Return of Material Instruction Manual April 2017 LIQ MAN 56 116 ...